KL-30B

Particle Counter, For pure water, Printer supplied

Monitors 0.05 μm particle contamination in pure water

  • Integrated unit combines sensor, controller, flow meter, and data storage
  • Install directly to pure water line and monitor the process
  • Purge air unit installed as standard
  • High pressure up to 500 kPa (gauge pressure) applicable
  • Equipped with D/A converter (4 - 20 mA) and alarm output contacts (standard)
  • Printer built-in, CompactFlash card available (option)

Specification

Optical system Light-scattering method
Light source Laser diode (wavelength 830nm, rated output 200mW )
Materials of parts exposed to sample fluid Synthetic quartz, fluorocarbon rubber, fluoroplastic,PVC, Pyrex glass, SUS304/316, POM
Measurement size range 4 channels (Factory default): ≥0.05µm, ≥0.1µm, ≥0.15µm, ≥0.2µm
Setting range 0.05µm to 0.2µm
Sanpling Flow rate Flow rate 10mL/min and purge flow rate 0.1 to 1L/min combined (Purge flow rate will differ depending on sample fluid pressure)
Maximum particle number concentration 200,000 particles/mL (coincidence loss 10%)
Power requirements 100V to 240V AC, 50/60Hz, Approx. 80VA
Dimensions Approx. 230 (H) × 330 (W) × 570 (D) mm
Weight Approx. 19.8kg

KL-30AX

Particle Counter, For pure water, Printer supplied

Detects particles down to 0.04μm size in pure water

  • Built-in leakage sensor triggers alarm output when internal leak detected
  • High pressure up to 500kPa (gauge pressure) applicable
  • Equipped with D/A converter (4-20mA) and alarm output contacts (standard)
  • Printer built-in, CompactFlash card available (option)

Specification

Optical system Light-scattering method
Light source Diode pumped solid state laser (wavelength: 532nm, rated output 500mW)
Materials of parts exposed to sample fluid Synthetic quartz, fluorocarbon rubber, fluoroplastic, PVC, SUS304/316, Pyrexglass, POM
Measurement size range 4 channels (Factory default): ≥0.04µm, ≥0.08µm, ≥0.1µm, ≥0.15µm
Setting range 0.04µm to 0.15µm
Sanpling Flow rate Flow rate 20mL/min and purge flow rate 0.1 to 1L/min combined (Purge flow rate will differ depending on sample fluid pressure)
Maximum particle number concentration 15,000 particles/mL (coincidence loss 10%)
Power requirements 100V to 240V AC, 50/60Hz, Approx. 130VA
Dimensions Approx. 230 (H) × 385 (W) × 570 (D) mm
Weight Approx. 24.8kg

KL-28BF

Particle Counter

Detects partciles down to 0.2μm size

  • Conpact and integrated use for faciliry monitoring system
  • Versatile array of interface: RS-232C, RS-485, 4-20mA, 1-6V
  • Multi-point monitoring is possible with optional software such as RP-monitor

Specification

Optical system Light-scattering method
Light source Laser diode (wavelength: 780nm, rated output 40mW)
Materials of parts exposed to sample fluid Sapphire, PFA, PTFE
Measurement size range 2 channels: ≥0.2µm, ≥0.5µm
Flow rate 10mL/min
Maximum particle number concentration 1,200 particles/mL (coincidence loss 5%)
Power requirements supplied by KE-28B, 100V AC, Approx. 23VA
Dimensions Approx. 70 (H) × 85 (W) × 118 (D) mm
Weight Approx. 0.6kg

KL-28B

Particle Counter

Detects partciles down to 0.2μm size

  • Conpact and integrated use for faciliry monitoring system
  • Versatile array of interface: RS-232C, RS-485, 4-20mA, 1-5V
  • Multi-point monitoring is possible with optional software such as RP-monitor

Specification

Optical system Light-scattering method
Light source Laser diode (wavelength: 780nm, rated output 40mW)
Materials of parts exposed to sample fluid Synthetic quartz, PFA, PTFE
Measurement size range 2 channels: ≥0.2µm, ≥0.5µm
Flow rate 10mL/min
Maximum particle number concentration 1,200 particles/mL (coincidence loss 5%)
Power requirements supplied by KE-28B, 100V AC, Approx. 23VA
Dimensions Approx. 70 (H) x 85 (W) x 118 (D) mm
Weight Approx. 0.6kg

KS-16F

Particle Sensor

Detects partciles down to 0.1μm size

  • Process monitoring at critical system points
  • Integrated use in central chemical supply systems and wet benches
  • Counting efficiency is 60% and higher
  • Deionized water, chemical solution, hydrofluoric (KS-16F) measurements
  • Display, control and printout are possible with KZ-70
  • Direct control by computer (using KZ-70 together is possible)
  • Build up real time monitoring system with the software of RP-monitor or KF-03

Specification

Optical system Light-scattering method
Light source Laser diode (wavelength: 830nm, rated output 200mW)
Materials of parts exposed to sample fluid Sapphire, PFA
Measurement size range 5 channels: ≥0.1µm, ≥0.15µm, ≥0.2µm, ≥0.3µm, ≥0.5µm
Flow rate 10mL/min
Maximum particle number concentration 1,200 particles/mL (coincidence loss 5%)
Power requirements 100V to 240V AC, Approx. 40VA (including external power requirement KZ-50 (accessory))
Dimensions Approx. 110 (H) x 240 (W) x 150 (D) mm
Weight Approx. 3.5kg

KS-16

Particle Sensor

Detects partciles down to 0.1μm size

  • Process monitoring at critical system points
  • Integrated use in central chemical supply systems and wet benches
  • Counting efficiency is 65% and higher
  • Display, control and printout are possible with KZ-70
  • Direct control by computer (using KZ-70 together is possible)
  • Build up real time monitoring system with the software of RP-monitor or KF-03

Specification

Optical system Light-scattering method
Light source Laser diode (wavelength: 830nm, rated output 200mW)
Materials of parts exposed to sample fluid Synthetic quartz, PFA
Measurement size range 5 channels: ≥0.1µm, ≥0.15µm, ≥0.2µm, ≥0.3µm, ≥0.5µm
Flow rate 10mL/min
Maximum particle number concentration 1,200 particles/mL (coincidence loss 5%)
Power requirements 100V to 240V AC, Approx. 40VA (including external power requirement KZ-50 (accessory))
Dimensions Approx. 110 (H) x 240 (W) x 150 (D) mm
Weight Approx. 3.5kg

KS-17B

Particle Sensor, For pure water

Detects particles down to 0.05μm size in pure water

  • Can be directly installed in equipment for on-site process control
  • Max. Sample pressure is 700 kPa
  • Sensor controller KZ-70 can be used for display, operation control, printout, and other functions
  • Multi-point monitoring system interface allows control form a computer (also in conjunction with KZ-70) directry
  • Allows configuration of multi-point monitoring system (RP monitor/KF-03)

Specification

Optical system Light-scattering method
Light source Laser diode (wavelength: 830nm, rated output 200mW)
Materials of parts exposed to sample fluid Synthetic quartz, PFA, Fluorocarbon rubber
Measurement size range 2 channels: ≥0.05µm, ≥0.1µm
(4 channels optional)
Flow rate 10mL/min
Maximum particle number concentration 100 000 particles/mL (coincidence loss 5%)
Power requirements 100V to 240V AC, Approx. 40VA (including external power requirement KZ-50 (accessory))
Dimensions Approx. 160 (H) x 300 (W) x 250 (D) mm
Weight Approx. 6kg

KL-30A

Particle Counter, For pure water, Printer supplied

Detects particles down to 0.05μm size in pure water

  • Integrated unit combines sensor, controller, flow meter,and data storage
  • Built-in leakage sensor triggers alarm output when internal leak detected
  • High pressure up to 500kPa (gauge pressure) applicable
  • Equipped with D/A converter (4-20mA) and alarm output contacts (standard)
  • Printer built-in, CompactFlash card available (option)

Specification

Optical system Light-scattering method
Light source Diode pumped solid state laser (wavelength: 532nm, rated output 500mW)
Materials of parts exposed to sample fluid Synthetic quartz, fluorocarbon rubber, fluoroplastic, PVC, SUS304/316, Pyrexglass, POM
Measurement size range 4 channels (Factory default): ≥0.05µm, ≥0.1µm, ≥0.15µm, ≥0.2µm
Setting range 0.05µm to 0.2µm
Flow rate Flow rate 20mL/min and purge flow rate 0.1 to 1L/min combined (Purge flow rate will differ depending on sample fluid pressure)
Maximum particle number concentration 15,000 particles/mL (coincidence loss 10%)
Power requirements 100V to 240V AC, 50/60Hz, Approx. 130VA
Dimensions Approx. 230 (H) x 385 (W) x 570 (D) mm
Weight Approx. 24.8kg