| KC-22A |
| (KC-22A is not available in U.S.) |
| 0.2 micron, 2.83 liter/min |
 |
Compact dimensions - Low Weight - High Output - Long Service Life
- Compact, lightweight, low-cost model with high reliability
- Newly developed light source uses laser diode excitation principle for optimum reliability and mechanical strength
- Long-life laser diode (manufacturer rating: MTBF 110.000 hours)
- Designed for easy maintenance and quick adjustment
- Standard RS-232C interface allows unattended measurement under computer control or data output to printer (KP-06, option)
|
| Specifications |
| Optical system |
90 degree sidway light-scattering method |
| Light source |
Diode-pumped solid state laser (wavelength 1064 nm), open-cavity type
Laser diode: wavelength 800 nm, rated output 1 W
Solid-state laser: Nd:YVO4 |
|
Particle size measurement ranges
|
5 channels (0.10, 0.15, 0.20, 0.30, 0.50 micron) |
| Flow rate |
2.83 liter/min |
| Concentration |
0 to 10,000particles/liter (Coincidence loss: 5% or less) |
| Dimensions, weight |
Approx. 185(H) X 155(W) X 330(D)mm (excluding protruding parts),
approx. 7.5 kg |
|
|
| KC-22B |
| (KC-22B is not available in U.S.) |
| 0.08 micron, 300mL/min |
 |
- High concentration for 0.08 micron measurement
- Precise measurement with small amount of sample air
- Compact and lightweight. High output with long life.
- RS-232-C interface enables autmoatic measurement by computer and printout
with the printer KP-06 (opton).
|
| Specifications |
| Optical system |
90 degree sidway light-scattering method |
| Light source |
Diode-pumped solid state laser (wavelength 1064 nm), open-cavity type
Laser diode: wavelength 800 nm, rated output 1 W
Solid-state laser: Nd:YVO4 |
| Particle size measurement ranges |
5 channels (0.08, 0.1, 0.2, 0.3, 0.5 micron and over) |
| Flow rate |
300 mL/min |
| Max. Concentration |
100,000 particles/liter (Coincidence loss: 5% or less) |
| Built-in Interface |
Serial interface (supplied)
D/A converter interface KZ-25L (factory option) |
| Dimensions, weight |
Approx. 155 (W) X 185 (H) X 330 (D) mm (excluding protruding parts), Approx.
7 kg |
|
|
| KC-20 |
| 10 micron to 100 micron, 30 liter/min |
 |
- For coarse particles measurement: from 10 to 100 micron
- Flow rate: 30 liter/min
- Alarm function
|
| Specifications |
| Optical system |
70 degree sideway light-scattering method |
| Light source |
Laser diode (780nm) |
| Particle size measurement ranges |
5 channels (10, 20, 30, 50, 100 micron) |
| Flow rate |
30 liter/min |
| Concentration |
0 to 2,000 particles/liter (Coincidence loss: 5% or less) |
| Dimensions, weight |
Approx. 100(H) X 290(W) X 405(D) mm, Approx. 12 kg |
|
|
| KC-01D1 |
| 0.3 micron, 0.5 liter/min |
 |
- Light-weighted but rigid body
- Automatic measurement and systematization with built-in RS-232C
- Unattended measurement by connecting optional printer
|
| Specifications |
| Optical system |
70 degree sidway light-scattering method |
| Light source |
Laser diode |
| Particle size measurement ranges |
5 channels (0.3, 0.5, 1, 2, 5 micron) |
| Flow rate |
0.5 liter/min |
| Concentration |
0 to 100,000 particles/liter (Coincidence loss: 5% or less) |
| Dimensions, weight |
Approx. 99(H) X 289(W) X 350(D) mm, Approx. 7 kg |
|
|
| KC-03A1 |
| 0.3 micron, 3 liter/min |
 |
- Counts particle in Class 100 clean room
- 4-step Alarm function can control air filtering equipment
- Optional built-in type interfaces (RS-232C, GP-IB, D/A converter) and optional
printer for measurement system
|
| Specifications |
| Optical system |
70 degree sideway light-scattering method |
| Light source |
Laser diode |
| Particle size measurement ranges |
5 channels (0.3, 0.5, 1, 2, 5 micron) |
| Flow rate |
3 liter/min |
| Concentration |
0 to 30,000 particles/liter (Coincidence loss: 5% or less) |
| Dimensions, weight |
Approx. 99(H) X 289(W) X 350(D) mm, Approx. 7.5 kg |
|
|
| KA-80B/KA-82 (KA-82: Soon-to-be-released) |
| (KA-82 is not available in U.S.) |
| 0.3 micron, 1liter/min |
 |
- For simultaneous, multi-point monitoring system
- Installation for monitor point is easily changeable
- Combined with pump, small sensors can be installed at every measuring point
- Sensors can be installed up to 20 measurement points with RP-monitor (software,
option)
- Multi-point monitoring system with 100 measurement points using KF-02A
(software, option)
|
| Specifications |
|
KA-80B |
KA-82 |
| Optical system |
70 degree sidway light-scattering method |
90 degree sidway light-scattering method |
| Light source |
Laser diode |
Open cavity type He-Ne laser diode |
| Size ranges |
2 channels (0.3, 0.5 micron) |
3 channels (0.1, 0.3, 0.5 micron) |
| Flow rate |
1 liter/min |
1 liter/min |
| Concentration |
60 000 particles/liter (Coincidence loss 5%) |
50 000 particles/liter (Coincidence loss 5%) |
| Dimensions |
Approx. 186(H) X 122(W) X 330 (D) mm |
Approx. 184(H) X 155(W) X 336(D)cm |
| Weight |
Approx. 4.5 kg |
Approx. 7 kg |
|
|
| KM-27 |
| 0.3 micron, 28.3 liter/min |
 |
- Simultaneous measurement in six ranges (from 0.3 to 10 micron)
- Large sample flow ensures fast and precise measurement and evaluation
- Built-in thermal printer
- Serial interface allows continuous measurement under computer control
- Buffer memory for 400 data
- Manifold type multi-tube system configuration possible (option)
|
| Specifications |
| Optical system |
70 degree sideway light-scattering method |
| Light source |
Laser diode |
| Particle size measurement ranges |
6 channels (0..3, 0.5, 1.0, 3, 5, 10 micron) |
| Flow rate |
28.3 liter/min (1CF/min) |
| Concentration |
14 000 particles/liter(Coincidence loss 5% for 0.3micron particle) |
| Dimensions, weight |
Approx. 285(W) X 156(H) X 448(D)mm(excluding protruding parts), Approx.
11 kg |
|
|