Particle Counter for Injections


Light Obscuration Automatic Particle Counter system KL-04 
Efficient measurement system for monitoring insoluble particle content in liquids and pharmaceutical injections
Standardized measurement system conforming to JP 14, USP24<788>, EP and BP.
  • Software allows automatic calibration
  • Sensor resolution accuracy tolerance within 10% according to USP 24 <788> method
  • Measuremet range 1.3 to 100 micron
  • Integrated hard disk can store the several years' measurement data
  • Data also can be stored in the integrated tape drive
  • Selectable display language: English, German, French, Japanese
  • Printer (option) can be connected for the printout of the data
Specifications
Optical system Light obscuration
Light source Laser diode (output 5mW, wavelength 780nm)
Materials of sampling tube Fused silica, PFA, Kalrez
Measurable particle size range 1.3 to 100 micron
Sample Flow rate 25 mL/min (10mL/min:option)
Max. particle concentration 10,000 particles/mL (coincidence loss 10% for 10 micron m PSL particles)
Counting efficiency 100 +/-5%(measuring PSL particles in the range of 10 micron, using count of 5 micron and above for comparison with reference unit)
Syringe capacity 25 mL (10mL: option)
Suction and drainage flow rate of syringe pump 5 to 10 mL/min
LAN connector Network interface (10Base-T) is built in.
Display 6 inch color LCD
Dimensions and weight Approx. 385(W) X 370(H) X 230(D) mm (excluding protruding parts.),
Approx. 15 kg