[Previouse] [Index] [Home]

Portable Particle Counter for Oil measurement
PODS -- Portable Oil Diagnostic System KR-60


On-site measurement of particles in OIL, viscosity and temperature of OIL
World first particle counter which is calibrated by new ISO-11171

Pods is an intelligent, portable and robust analysis instrument for measuring, storing and reporting of oil condition parameters important for reliable hydraulic system operation.

PODS provides the means to analyze pressurize fluids and lubricants online or bottle sampling mode without interrupting machine operations.
KR-60写真
PODS measures three oil parameters:
1. Particle Concentration (from which cleanliness classes based on several standards are derived)
2. Oil Temperature
3. Oil Viscosity
PODS has three operating modes:
1. Analysis from bottle samples
2. Online measurement at the pressure of the hydraulic system up to 420 bar or 6000 psi.
3. Online measurements in combination with an ARGO ECOTEC "Oil Service and Transfer Cart".
For example during oil exchange and refilling or off-line filtration of a hydraulic system.
Specifications
Detected method Light extinction
Light source Laser diode
Size range 4 to 68micron m(ISO-MTD)、 1 to 100micron m(ACFTD)
Size channels 8 channels
Channel 1 2 3 4 5 6 7 8
ISO-MTD [micron] 4 4.6 6 9.8 14 21.2 38 68
ACFTD [micron] to 1 2 to 5 10 to 15 25 to 50 to 100
Flow rate 20 to 50 mL/min (automatic control with feedback circuit)
Calibration Hydraulic oil for airplane MIL H-5606
ISO-MTD(ISO 11171:1999)
Cleanliness class ISO 4406, NAS 1638, MIL-STD-1246C, NAVAIR 01-1A-17
Cleanliness class range ISO 4406 code 4 to 23
Max. particle concentration 90 000particles/mL (counting loss 10%)
Display of measurement results Cleanliness class (by each standards), Concentration (particles/mL), Temperature (C), Viscosity (mm2/S)
Parts materials exposed to sample oil Stainless steel, sapphire, aluminum, PTFE
Sample oil pressure 7 to 420 bar
Sample oil temperature 0 to 90C
Sample oil viscosity 10 to 430mm2/s (When flow rate error is about 50%.)
Operation modes Bottle sampling mode: max. 7 bar
On-line sampling mode: max. 420 bar
Pressure source to supply with fluid Supplied fluid CO2(Measurable about 50 times), External pressurized air (7 bar or less) or Pressure from the unit (online measurement)
Sample bottle Supplied 100-mL bottle
Measurement operation After 15-mL purging for each sample is done, 3-cycle measurements for each sample by 10 mL per one cycle are carried out. Measurement results are stored in buffer, or printed out.
Buffer memory capacity 500 data
RS-232C interface port Digital output 0 to 5Vdc/<20 mA no loaded, analog input 0 to 5 Vdc
Operation mode Batch sampling (7bar), online sampling (420 bar)
Software (option) "PODSWare" (OS: Windows 95/98/NT)
Download of memory and online mode are possible.
Power supply AC100 to 240 V, 50/60 Hz
Dimensions and weight 300 X 350 X 200 mm, Approx. 11.5 kg

OPTION
Software <PODS Ware> Buffer memory download, online remote control mode

Portable recirculation filtration unit UMP-45P Connected with hydraulic system to recirculate and filtrate the contaminated oil.

Connected with PODS to recirculate and filtrate the oil while monitoring the cleanliness of the oil. Automatic control is possible.

Flow rate 25 to 45L/min
Filter ability 1micron m abs. 265g ACFTD
3micron m abs. 350g ACFTD
10micron m abs. 400g ACFTD



[Previouse] [Index] [Home]